How can we help?

Search and find answers to frequently asked questions.

If a user of the twiist AID system temporarily loses connection with the FreeStyle Libre 3 Plus sensor, will the missing data be backfilled?

Yes. When the twiist system has not received a CGM reading from the FreeStyle Libre 3 Plus sensor in 5 minutes, CGM Signal Loss will be displayed in the CGM Status.

When the twiist system has not received a CGM reading in 20 minutes or more, a Signal Loss CGM alert will be displayed. The last 24 hours of lost glucose data will be automatically retrieved when the connection is restored. The twiist User Guide recommends wearing the sensor and pump on the same side of the body to minimize the risk of losing Bluetooth connectivity that can cause data gaps.

Sequel: the twiist™ AID System
For full twiist™ AID system safety information, please visit www.twiist.com/safety.


Was this information helpful?

Thank you for your feedback.

We want to help you receive the information you need. Other options you might consider:

Contact Customer Care

1-855-632-8658

Available 7 days a week 8 AM to 8 PM Eastern Time; excluding Holidays

Go to FAQ Search

ADC-2661293 v5.0 06/26