What is MARD?
Another evaluation method used in clinical studies to represent device accuracy data. The MARD (Mean Absolute Relative Difference) measures the average difference between a device measurement (or test result) and the reference measurement at normal to high glucose levels. The lower the MARD, the better the agreement between the device and the reference/comparator measurement. Excellent accuracy on the first day (MARD is 11.2% for adults and 12.2% for pediatrics) - Sensor readings closer to blood glucose readings¹ - Even better overall accuracy for the entire wear period (99.9% of readings in Zones A and B and MARD is 9.2% for adults, and 100% of readings in Zones A and B and MARD is 9.7% for pediatrics). The accuracy was achieved without finger prick calibration or any calibration code entry.*
*Data on File, Abbott Diabetes Care Inc, FreeStyle Libre 2 System Performance Data Sheet.